Will dental implants set off airport scanners, and can I have an MRI?
Almost never at the airport, and yes to the MRI. A dental implant is a small titanium screw, and in an airport study people whose only metal was four or fewer screws never set off the walk-through metal detector (Obremskey 2007). Australian airports now use millimetre-wave body scanners, whose waves reflect off the skin and do not pass into the body, so they do not see implants inside your jaw (Melbourne Airport). For MRI, titanium is not magnetic. Lab tests found implants warmed by less than half a degree in a 1.5 tesla scanner (Paknahad 2024), and implant bridges warmed by under 2°C at 3 tesla, which the authors judged no risk (Hasegawa 2013). The real effect is on the picture: titanium blurs the image a few millimetres around each implant (Bohner 2020), which matters only for scans of the jaw, face or nearby. Tell the radiographer you have implants, and say so if you have a denture held on by magnets, because magnetic parts need checking ahead of the scan and again once it is over (Hasegawa 2015).
Implants, scanners and MRI, at a glance
| Situation | What to expect | What to do |
|---|---|---|
| Airport body scanner (millimetre wave) | It does not see inside the body, so it does not show implants | Nothing; tell the officer if you are asked about medical devices |
| Walk-through metal detector | Very unlikely to alarm for dental implants alone | If it alarms, say you have dental implants |
| Handheld wand close to the face | Not well studied; a metal bar in a full-arch bridge may register at close range | Say you have dental implants; show your implant card if asked |
| MRI of the body, away from the head | No effect on the scan | Tell the radiographer you have dental implants |
| MRI of the head, face, jaw or neck | Safe; the image may be blurred a few millimetres around each implant | Tell the radiographer, so they can adjust the scan |
| MRI with a denture held on by magnets | The magnet parts are pulled by the scanner | Tell the radiographer; take the denture out; have the magnet parts checked |
| CT scan of the head | Metal causes streaks on the image near the implants | Tell the radiographer |
The short version: implants are a non-event at the airport, and safe in an MRI. The one thing worth remembering is to tell the person doing the scan.
What an implant is made of, and why it matters
A dental implant is a small screw placed in the jawbone to act as a tooth root. Most are made of titanium or a titanium alloy. A few are made of zirconia, a white ceramic with no metal. On top of the implant sits a connector called an abutment, often titanium too, and then a crown.
- Size: a single implant is a few millimetres wide and around a centimetre long, far smaller than a hip or knee joint.
- Titanium: a metal, so a metal detector can sense it in principle, but it is not magnetic.
- Zirconia implants and crowns: ceramic, with no metal at all.
- All-on-4: four or more implants, often joined by a bridge with a metal frame or bar inside, so more metal in one place.
- Bone graft screws and pins: some grafts are held with tiny titanium screws, which add very little metal.
Two different things are tested at an airport and in an MRI. A metal detector senses any metal that conducts electricity. An MRI is a strong magnet, so what matters there is whether a metal is magnetic. Titanium conducts, but it is not magnetic.
Airport security: body scanners
Australian airports screen passengers mainly with body scanners. These use millimetre waves, a kind of radio wave, not X-rays. Melbourne Airport states that all body scanners in Australia use this technology, that the waves reflect off the surface of the skin and do not pass into the body, and that the image shows no human anatomy (Melbourne Airport).
That means a body scanner looks for items worn or carried on you, not for what is inside you. An implant in your jaw sits under gum and bone. The scanner does not see it.
- You do not need a letter for dental implants to go through a body scanner.
- Melbourne Airport asks passengers to tell staff about medical devices, aids or implants before screening. Saying "I have dental implants" is enough if you are asked.
- If you want to decline the body scanner for a medical reason, Melbourne Airport asks for a letter from a medical practitioner. Dental implants are not a reason to decline it.
Airport security: walk-through metal detectors and wands
Some airports, and some passengers who cannot use a body scanner, still go through walk-through metal detectors. No study has tested dental implants in them directly, but orthopaedic implants give a useful comparison.
In a US trauma clinic, 96 patients with orthopaedic implants walked through an airport arch detector set to standard sensitivity (Obremskey 2007). People whose only metal was four or fewer screws were never detected, 0 of 7. Every joint replacement was detected. Long plates and titanium rods were the other things that set it off (Obremskey 2007).
| Implant | Detected by the arch | How it compares with dental implants |
|---|---|---|
| Four or fewer screws only | Never (0 of 7) | Closest to dental implants in amount of metal |
| Long plates (more than 10 holes) | Often | Far more metal than any dental work |
| Titanium rods in long bones | Often | Far more metal than any dental work |
| Hip or knee replacement | Always in this study | Far more metal than any dental work |
Single implants and small bridges hold less metal than a few bone screws. A full-arch bridge with a metal bar holds more, but still far less than a joint replacement. A handheld wand is held close to the body, and how it reacts to a full-arch bridge has not been studied.
Even with joint replacements, most people get through with a short explanation. In a UK survey of 50 people with hip or knee replacements, 14% said their joints had never set off the alarm, and 90% of the airport officers asked said an implant card was useful to them (Ali 2017).
What to carry and say at the airport
- Keep the implant card or record from your clinic with your passport. It lists the brand and the teeth treated, and it is useful at the dentist at home too.
- If an officer asks about medical devices, say you have dental implants in your jaw.
- If an alarm goes off, say the same and point to your mouth. A short pat-down or wand check usually follows.
- If you are flying home straight after implant surgery, carry your aftercare papers and medicines in your hand luggage, with the clinic's number.
Security is rarely the issue on the flight home. Swelling, pain relief and when it is safe to fly after surgery matter more, and the clinic sets your flight date with those in mind.
MRI: is it safe with dental implants?
Yes. An MRI scanner uses a powerful magnet and radio waves to make pictures. Three things could in theory go wrong with metal in the body: it could be pulled by the magnet, it could heat up, or it could spoil the picture. For titanium dental implants, the first two are not a concern in the studies.
| Concern | What the studies found | Source |
|---|---|---|
| Pulled by the magnet | Titanium is not magnetic, so implants are not pulled | Basic property of titanium |
| Heating at 1.5 tesla | Titanium implants warmed by 0.35 to 0.47°C; judged MR safe for heating | Paknahad 2024 |
| Heating at 3 tesla | Metal bridges warmed by up to 1.80°C and a full-arch bridge by 1.59°C; judged no risk | Hasegawa 2013 |
| Blurring of the picture, titanium | About 2.57 mm of blur around titanium implants | Bohner 2020 |
| Blurring of the picture, zirconia | About 0.37 mm around zirconia implants | Bohner 2020 |
The two strengths tested above, 1.5 and 3 tesla, are the ones most hospital scanners use. The temperature rises found are small, well under the warmth of a hot drink in the mouth.
The radiographer will still ask about metal in your body. That is routine, and the answer helps them, so always mention dental implants, bridges, crowns with metal and any magnets in a denture.
MRI: what implants do to the picture
Metal disturbs the magnetic field right next to it, which shows as a dark blur or distortion on the image. Doctors call this an artefact. With dental implants the blur is local.
- Scans away from the head, such as the knee, back or abdomen: dental implants have no effect.
- Brain scans: often readable, because the implants sit low in the face. The radiographer may adjust the settings.
- Scans of the jaw, tongue, salivary glands, sinuses or neck: the blur near the implants can hide detail, and the radiographer may adjust the scan or choose a different test.
Material makes a difference. In a test in pig bone at 3 tesla, titanium implants caused about 2.57 mm of blur around them and zirconia implants about 0.37 mm (Bohner 2020). Another lab study found zirconia caused only minor distortion on MRI while titanium caused extensive artefacts; on CT and on the dental 3D scan the reverse was true (Demirturk Kocasarac 2019).
Dental materials in general cause more trouble on CT than on MRI. In a test of 44 dental materials, 13 caused artefacts on MRI and 41 caused artefacts on CT (Klinke 2012). Crowns with metal inside, old amalgam fillings and metal bars all add to it.
Dentures held on by magnets: the exception
Some implant dentures clip on with small magnets: a magnet in the denture and a steel keeper fixed to each implant. Steel keepers are magnetic, so they behave differently in an MRI.
- Heating stayed below 1.0°C at 3 tesla in a test of magnetic attachments, so heat is not the issue (Miyata 2012).
- The pull is the issue: the keepers were pulled strongly by the scanner, so they must be firmly fixed before a scan (Miyata 2012).
- Inspect magnetic attachments ahead of an MRI and again afterwards, and remove large prostheses with cast magnetic alloy (Hasegawa 2015).
- The removable denture itself, with its magnets, comes out before the scan, like any removable denture.
- Magnets near the face also blur the picture more than titanium does.
All-on-4 and other fixed bridges screwed to implants do not use magnets. Many snap-on dentures use clips called locators rather than magnets. If you are not sure which you have, your implant record or your dentist can tell you.
Your situation, and what to do
| Your dental work | Airport | MRI |
|---|---|---|
| One or a few titanium implants with crowns | No action needed | Tell the radiographer; no other change |
| Zirconia implants or metal-free crowns | No action needed | Tell the radiographer; little blur |
| All-on-4 or All-on-6 with a metal frame | Say you have dental implants if an alarm or a wand registers | Tell the radiographer; expect blur near the jaw on head scans |
| Snap-on denture with locator clips | No action needed | Take the denture out; tell the radiographer |
| Denture held by magnets | No action needed | Tell the radiographer; take the denture out; have the keepers checked ahead of the scan and afterwards |
| Implant surgery days before your flight | Carry aftercare papers and medicines in hand luggage | Tell the radiographer about the recent surgery |
What this means if you fly from Australia to Vietnam
- Implant treatment means two trips of about 7 days each, 3 to 6 months apart while the bone heals. You go through airport security at least four times. Implants are not a problem at any of them.
- Keep the implant card or written record from the clinic, with the brand, size and tooth numbers. It helps at the airport if ever needed, and it helps any dentist or radiographer later.
- If you have an MRI booked at home, tell the imaging centre you have dental implants when you book, not only on the day.
- If you might need regular head or neck MRI for another condition, mention it before treatment. The dentist can weigh zirconia or titanium for you.
For how long implants last and what can go wrong over the years, see how long dental implants last.
Common beliefs that are not quite right
- "Implants will set off the airport alarm every time." Small amounts of metal, like a few screws, did not set off an arch detector in the study above (Obremskey 2007), and body scanners do not see inside you.
- "I need a doctor's letter to fly with implants." Not for dental implants. A letter is only needed if you want to skip the body scanner for a medical reason.
- "You cannot have an MRI with implants." You can. Titanium is not magnetic, and the heating in tests was small (Paknahad 2024; Hasegawa 2013).
- "An MRI can pull an implant out of the bone." Titanium is not pulled by the magnet. Magnetic denture keepers are the exception, and they are checked before the scan.
- "Zirconia implants are needed for MRI." Not for safety. Zirconia only reduces the blur on scans near the jaw.
Where the evidence is thin
No published study has walked people with dental implants through airport metal detectors, so the airport advice rests on orthopaedic studies and the size of the metal. MRI heating studies were done in the lab, on phantoms, extracted teeth or animal bone, not in people. Their results agree with each other and with years of routine scanning.
Quick answers
Do dental implants show up on airport body scanners?
No. Australian body scanners use millimetre waves that reflect off the skin and do not pass into the body, so implants in your jaw are not seen.
Can I have a 3 tesla MRI with dental implants?
Yes. In lab tests at 3 tesla, metal dental bridges warmed by less than 2°C and the authors judged them no risk (Hasegawa 2013). Tell the radiographer you have implants.
Will my implants ruin an MRI of my head?
Usually not. Titanium blurs the picture a few millimetres around each implant (Bohner 2020). Brain scans are often readable; scans of the jaw or mouth may need adjusted settings.
Should I carry an implant card when I travel?
It is useful. Keep the record of the brand, size and teeth with your passport. Airport officers say implant cards help (Ali 2017), and any dentist at home will need the details.
Are zirconia implants better for MRI?
They blur the MRI picture less than titanium (Bohner 2020). Both are safe in an MRI, so this matters only if you need regular scans of the face or jaw.
Questions about implants and travel?
Tell us your plans and any scans you have coming up. We plan the trips around your flights and make sure you leave with the full record of your implants. Free.
Talk to usSources
- Obremskey WT et al. Detection of orthopaedic implants by airport metal detectors. J Orthop Trauma, 2007.
- Ali E, Kosuge D, MacDowell A. The need for an implant identification card at airport security check. Clin Orthop Surg, 2017.
- Melbourne Airport. Security screening (body scanners, medical devices and implants), accessed November 2026.
- Paknahad M, Khaleghi I, Mortazavi SMJ. Radiofrequency-induced heating of amalgam restorations and dental implants during 1.5T magnetic resonance imaging. Front Dent, 2024.
- Hasegawa M et al. Radiofrequency heating of metallic dental devices during 3.0 T MRI. Dentomaxillofac Radiol, 2013.
- Bohner L et al. Magnetic resonance imaging artifacts produced by dental implants with different geometries. Dentomaxillofac Radiol, 2020.
- Demirturk Kocasarac H et al. Evaluation of artifacts generated by titanium, zirconium, and titanium-zirconium alloy dental implants on MRI, CT, and CBCT images: a phantom study. Oral Surg Oral Med Oral Pathol Oral Radiol, 2019.
- Klinke T et al. Artifacts in magnetic resonance imaging and computed tomography caused by dental materials. PLoS One, 2012.
- Miyata K et al. Radiofrequency heating and magnetically induced displacement of dental magnetic attachments during 3.0 T MRI. Dentomaxillofac Radiol, 2012.
- Hasegawa M et al. 3-T MRI safety assessments of magnetic dental attachments and castable magnetic alloys. Dentomaxillofac Radiol, 2015.